Probe Numbering

Technology

• PEC: Single-element pulsed eddy current
• PECA: Pulsed eddy current array

Model Size

• 025: PEC Small
• 089: PEC Medium
• 152: PEC Large
• SM: PECA Small
• MED: PECA Medium

Model (application specific)

• HR: High-Resolution
• GS: Galvanized Steel
• SZ: Splashzone
• TF: Tank Floor
• UW: Underwater

Channels

• xCH: Number of PECA elements

Cable Type

• H: Heavy-duty Black Cable
• U: Underwater Neon Green cable

Cable Length

Indicates the length of the cable in meters.

Cable Exit Position

• S: Side Exit

 

Lyft Probe Numbering System

Pulse Eddy Current Array Probes

The PECA probes are specifically engineered for high productivity inspections in grid,
or high-resolution dynamic modes. The wide coverage not only makes inspections faster than ever,
but also improves data quality and minimizes inspection preparation with Eddyfi Grid-As-U-Go.

PECA-6CH-MED-H05S
Casing:Medium
Number of channels:6
Total coverage:457mm (18in)
Wall thickness:6-25mm (0.25-1.00in)
Footprint:46mm (1.80in)
Liftoff:Up to 76mm (3in)
Encoder:PEC-ENC-STD
Weather jacket:Stainless steel up to 1.5 mm (0.06 in)
Aluminum up to 1.0 mm (0.04 in)
Hardware Compatibility:Compatible with LYFT-PEC-GDA
Minimal software requirements:Lyft-Go 2.0
PECA 6CH Array Probe
PECA Probe Wall Thickness & Insulation Chart

Single-Element Pulse Eddy Current Probes

Single-element PEC probes are designed to detect CUI, CUF, and FAC in pipes, vessels,
sphere legs, small pipes and in limited access areas. They are available in three different sizes
for the right balance between wall thickness and liftoff. They offer enough flexibility
to support metals as thick as 100 mm (4 in), insulation up to 300 mm (12 in) thick
(fiber glass, plastic wrap, concrete, and any non-ferrous material), as well as
stainless steel, aluminum, or galvanized steel weather jackets.

PEC-025-G2-H05S
Casing:Small
Number of channels:1 channel (Single-element)
Wall thickness:3-19mm (0.12-0.75in)
Footprint:35mm (1.38in)
Liftoff:Up to 76mm (3in)
Encoder:PEC-ENC-STD
Weather jacket:Stainless steel up to 1.5 mm (0.06 in)
Aluminum up to 1.0 mm (0.04 in)
Galvan. steel up to 1.0 mm (0.04 in)
Hardware Compatibility:Compatible with all models
Minimal software requirements:Lyft-Go 1.2+
PEC-025-G2-H05S
PEC-089-G2-H05S
Casing:Medium
Number of channels:1 channel (Single-element)
Wall thickness:3-76mm (0.12-3.00in)
Footprint:62mm (2.44in)
Liftoff:Up to 203mm (8in)
Encoder:PEC-ENC-STD
Weather jacket:Stainless steel up to 1.5 mm (0.06 in)
Aluminum up to 1.0 mm (0.04 in)
Galvan. steel up to 1.0 mm (0.04 in)
Hardware Compatibility:Compatible with all models
Minimal software requirements:Lyft-Go 1.2+
PEC-089-G2-H05S
PEC Probe Wall Thickness & Insulation Chart
Close

Enquiry about Eddyfi LYFT Probes